Please select your Region.
Please select your Region.
The ADAMS A1c, HA-8190V is a fully automated Glycohemoglobin (HbA1c)
analyzer based on HPLC (High Performance Liquid Chromatography).
HA-8190V detects and separates variant hemoglobin automatically, and this function will be the new standard for hemoglobin analyzer.
Measurement Speed
User friendly
Better HbA1c results
Sample Type | Whole blood or hemolysis sample |
---|---|
Measurement items | HbA1c (stable HbA1c), HbF (In Variant mode, HbS and HbC can be separated and HbE and HbD can be detected) |
Measurement principle | HPLC method (Reversed-phase cation exchange chromatography) |
Measurement wavelengths |
420nm/500nm (Dual-wavelength colorimetry) |
Resolution | 0.1% Ratio. 1mmol/mol |
Sample consumption | Approx. 8μL (whole blood) |
Processing speed | Fast mode : 24 seconds / sample Variant mode : 58 seconds / sample |
Required sample volume |
Sample tube : 10mm or above from the bottom of sample tube Sample cup : 400μL or above |
Sample container | Sample tube : (φ13 or 15) x (75 - 100mm) Sample cup : 500μL |
Displayed range | HbA1c : 3 to 20 %, 9 to 195 mmol/mol HbF : 0.0 to 99.9% |
Authenticated-compatibility range | HbA1c : 4 to 16 %, 20 to 151 mmol/mol |
Number of samples loaded | One-way transportation : Maximum of 50 samples Loop transportation : Maximum of 100 samples *Maximum of 200 samples can be loaded using the optional"Side Sampler". |
Display | Color LCD (with touch screen) |
Built-in printer | Thermal printer, 58 mm thermal paper |
Memory capacity | 900 test results (including calibration results) |
External output | Serial 1 port (Can be optionally used as an Ethernet port) |
Measurement conditions | Temperature : 10 - 30 ℃ Humidity : 20 - 80%RH (No condensation) |
Power consumption | 300VA |
Dimensions | 530(W)x530(D)x530(H)mm * Not including the projection portions, eluent packs and hemolysis washing solution bottle |
Weight | Analyzer : Approx. 41kg Sampler unit : Approx. 4kg |
AG220425-HA8190V-EN